    C-Terminal Sequencing: 8 Critical Mistakes to Avoid in LC-MS/MS Analysis

      LC-MS/MS analysis is extensively applied in proteomics for protein identification, post-translational modification profiling, and quantitative studies. However, during C-terminal sequencing, researchers often make avoidable mistakes in experimental design, data acquisition, and data analysis, which can compromise data quality and the reliability of conclusions. This article highlights 8 critical mistakes commonly encountered in LC-MS/MS-based C-terminal sequencing, and provides practical strategies to avoid them, thereby enhancing the accuracy of proteomic results.

       

      Common Pitfalls in the Sample Preparation Stage

      Mistake 1: Inadequate Sample Preparation

      In C-terminal sequencing, sample purity critically influences downstream enzymatic digestion and mass spectrometry performance. Contaminants such as high salt concentrations, detergents, and host-cell proteins can disrupt ionization efficiency and reduce signal-to-noise ratios.

       

      Recommendations:

      (1) Remove impurities using ultrafiltration or gel filtration;

      (2) Maintain appropriate protein concentrations;

      (3) Store samples at –80°C to prevent repeated freeze-thaw cycles, which may lead to protein degradation or loss of C-terminal modifications.

       

      Technical Oversights in Digestion Strategies

      Mistake 2: Exclusive Reliance on Trypsin Digestion

      Trypsin exhibits limited cleavage specificity near C-terminal regions, potentially resulting in missed identification of critical peptides.

       

      Recommendations:

      (1) Combine trypsin with proteases of distinct specificity (e.g., Glu-C, Asp-N, Lys-C);

      (2) Consider non-specific proteases like proteinase K to improve sequence coverage;

      (3) Optimize digestion parameters including time, pH, temperature, and enzyme-to-substrate ratio to prevent under- or over-digestion.

       

      Hidden Errors in Mass Spectrometry Settings

      Mistake 3: Unoptimized Instrument Parameters

      Failure to adjust fragmentation mode, resolution, or normalized collision energy (NCE) based on sample characteristics can result in loss of C-terminal signals.

       

      Recommendations:

      (1) For post-translationally modified proteins, prioritize ETD or EThcD fragmentation;

      (2) Use HCD to enhance y-ion intensity, aiding C-terminal sequencing;

      (3) Calibrate the mass window and NCE through pilot experiments to ensure optimal performance.

       

      Inappropriate Liquid Chromatography Conditions

      Mistake 4: Improper Retention Time Settings

      C-terminal peptides often display distinct physicochemical properties from internal peptides. Poor separation or premature elution can lead to weak signals and overlapping peaks.

       

      Recommendations:

      (1) Use UHPLC systems to enhance chromatographic resolution;

      (2) Optimize the organic solvent ratio and pH in the mobile phase;

      (3) Select suitable columns (e.g., C18, C8) that match the hydrophobicity/hydrophilicity profile of C-terminal peptides.

       

      Configuration Errors in Data Analysis Pipelines

      Mistake 5: Inappropriate Database Search Parameters

      Default search settings often overlook non-specific cleavages and common C-terminal modifications, resulting in missed peptide identifications.

       

      Recommendations:

      (1) Enable non-specific enzyme cleavage in database search settings;

      (2) Include variable modifications such as C-terminal acetylation, amidation, and phosphorylation;

      (3) Maintain a stringent false discovery rate (FDR ≤ 1%) to ensure reliable identifications.

       

      Unbalanced Filtering Criteria

      Mistake 6: Overly Stringent or Permissive Filtering

      Excessively stringent filters (e.g., ultra-low FDR thresholds) may exclude genuine peptides, while overly relaxed criteria increase false positives.

       

      Recommendations:

      (1) Evaluate filtering parameters such as matching scores, peptide coverage, and modification sites;

      (2) Apply multi-metric scoring strategies to refine peptide selection.

       

      Overlooking C-terminal Modifications

      Mistake 7: Neglecting C-terminal Post-Translational Modifications

      Modifications such as acylation, ubiquitination, and glycosylation at the C-terminus can significantly alter fragmentation patterns and lead to failed identification.

       

      Recommendations:

      (1) Customize database search settings to include relevant C-terminal modifications based on the target protein;

      (2) Use PRM-based targeted validation to confirm ambiguous modification sites where necessary.

       

      Lack of Result Validation

      Mistake 8: Relying Solely on a Single Sequencing Run

      Drawing conclusions from a single experiment risks misinterpretation and false positives.

       

      Recommendations:

      (1) Validate results with alternative digestion strategies;

      (2) Use synthetic peptides to confirm key findings;

      (3) Perform site-directed mutagenesis or fusion-based assays to validate C-terminal sequences of interest.

       

      Accurate implementation of C-terminal sequencing via LC-MS/MS requires a thorough understanding of its distinctions from conventional proteomics workflows. Optimizing the entire pipeline—from sample preparation to data interpretation—demands a tailored approach that accounts for the unique chemical characteristics of protein carboxyl termini. MtoZ Biolabs offers professional N- and C-terminal sequencing services to support your research goals, helping you resolve technical challenges and accelerate your project with high-quality analytical solutions.

       

      MtoZ Biolabs, an integrated chromatography and mass spectrometry (MS) services provider. 

      Related Services

      C-Terminal Sequencing Service

    Submit Inquiry
    Name *
    Email Address *
    Phone Number
    Inquiry Project
    Project Description *

     

    How to order?


    /assets/images/icon/icon-message.png

    Submit Inquiry

    /assets/images/icon/icon-return.png