• Home
  • Biopharmaceutical Research Services
  • Multi-Omics Services
  • Support
  • /assets/images/icon/icon-email-2.png

    Email:

    info@MtoZ-Biolabs.com

    Mechanism of Mass Spectrometry-Based Peptide Identification

      Mass spectrometry (MS), a highly sensitive and specific analytical tool, has found widespread applications in proteomics research. Particularly, MS-based peptide identification technology enables the efficient and accurate detection of peptides in biological samples, providing critical data for protein quantification and functional studies.

       

      Sample Preparation and Peptide Generation

      Before MS analysis, the sample typically undergoes several pre-treatment steps. Proteins are first extracted from biological samples, followed by chemical or enzymatic digestion into smaller peptide fragments. Trypsin is one of the most commonly used enzymes, cleaving proteins at lysine and arginine residues to generate peptide fragments suitable for mass spectrometric analysis.

       

      Working Principle of Mass Spectrometry

      Peptide identification through MS relies on the core principles of the instrument, which consists of three primary components: an ion source, a mass analyzer, and a detector.

       

      1. Ion Source

      The first step in MS analysis is peptide ionization. The most common ionization techniques are Electrospray Ionization (ESI) and Matrix-Assisted Laser Desorption/Ionization (MALDI). These methods convert peptides into charged ions, making them detectable by the mass spectrometer.

       

      In ESI, the liquid sample is sprayed into fine droplets under high pressure, and as the solvent evaporates, charged peptide ions are left behind. MALDI, on the other hand, involves mixing the sample with a matrix and using laser energy to desorb and ionize the peptides. Both ionization techniques aim to generate charged peptides suitable for further analysis.

       

      2. Mass Analyzer

      After ionization, the charged peptide ions are directed into the mass analyzer, where they are separated based on their mass-to-charge ratio (m/z). Common types of mass analyzers include Time-of-Flight (TOF), quadrupole, and ion trap analyzers.

       

      (1) TOF Mass Spectrometers

      calculate the m/z ratio by measuring the time it takes for ions to travel through a field-free region, with lighter ions reaching the detector faster than heavier ions.

       

      (2) Quadrupole Mass Spectrometers

      use alternating electric fields to separate ions. By adjusting the frequency and voltage of the electric field, quadrupoles allow only ions within a specific m/z range to pass through.

       

      (3) Ion Trap Mass Spectrometers

      trap and accumulate ions before releasing them sequentially according to their m/z values for detection.

       

      3. Detector

      Once the ions are separated by the mass analyzer, they reach the detector, which records the number of ions and their corresponding m/z values, generating a mass spectrum. Each peak in the spectrum represents a charged peptide ion, with the peak position corresponding to its m/z value and the intensity indicating its abundance.

       

      Peptide Fragmentation and MS/MS Analysis

      For further peptide sequence identification, MS is often coupled with tandem mass spectrometry (MS/MS). In MS/MS, peptides are first subjected to an initial mass analysis (MS1), where a specific precursor ion is selected and induced to fragment through Collision-Induced Dissociation (CID). The resulting fragments are then analyzed in a second stage of mass spectrometry (MS2), producing a secondary mass spectrum. By analyzing the m/z values of these fragments, the precursor peptide's amino acid sequence can be inferred.

       

      During CID, charged peptides typically break along the peptide bond, producing two main types of ions: b-ions and y-ions. b-ions are positively charged fragments originating from the N-terminus, while y-ions originate from the C-terminus. The m/z values of these ions allow researchers to deduce the peptide’s amino acid sequence.

       

      Data Analysis and Peptide Identification

      Once the mass spectra are generated, peptide identification relies on specialized software tools and database searches. Common databases like SwissProt and UniProt contain vast amounts of known protein sequences. MS software compares the experimental mass spectra with theoretical peptide spectra in these databases to match the closest sequences.

       

      Popular software tools include Mascot, Sequest, and X!Tandem, which calculate theoretical m/z values for peptide fragments and match them with experimental data to identify peptide sequences. Researchers may also use multiple search algorithms and statistical methods like False Discovery Rate (FDR) filtering to optimize and validate results.

       

      Applications and Challenges in Peptide Identification

      Mass spectrometry-based peptide identification is widely applied in proteomics, biomarker discovery, and drug target identification. However, challenges such as sample complexity, peptide charge states, and variations in ionization efficiency can affect detection and identification accuracy, necessitating ongoing improvements in MS technology and data analysis methods.

       

      The mechanism of mass spectrometry-based peptide identification offers powerful tools for protein research, allowing precise peptide sequence identification through ionization, mass analysis, fragmentation, and data matching.

    Submit Inquiry
    Name *
    Email Address *
    Phone Number
    Inquiry Project
    Project Description *

     

    How to order?


    /assets/images/icon/icon-message.png

    Submit Inquiry

    /assets/images/icon/icon-return.png