Workflow of Protein Identification by Tandem Ms
Tandem mass spectrometry (MS/MS) is a critical technology in modern proteomics research. Known for its high sensitivity and specificity, it is widely used for protein identification and quantification. This article will detail the workflow of protein identification by tandem mass spectrometry, including the main steps of sample preparation, mass spectrometry analysis, and data processing.
Sample Preparation
Sample preparation is the first step in protein mass spectrometry analysis and directly affects the quality of the mass spectrometry data and subsequent analysis results. Sample preparation usually includes the following steps:
1. Protein Extraction
Extract total proteins from biological samples (such as cells, tissues, or body fluids). This process typically includes cell lysis, protein precipitation, and purification. Common methods for cell lysis include mechanical disruption, sonication, and chemical lysis.
2. Protein Quantification
Quantify the extracted proteins using methods such as the Bradford assay, BCA assay, or UV absorbance to ensure consistent protein concentration in each sample.
3. Protein Digestion
Digest the extracted protein samples, usually using trypsin, to break down proteins into smaller peptides. This step is conducted under appropriate temperature and buffer conditions, typically incubating at 37°C for 16 hours.
4. Peptide Purification
Purify the digested peptides to remove undigested proteins, salts, and other interfering substances. Common purification methods include solid-phase extraction (SPE) and high-performance liquid chromatography (HPLC).
Mass Spectrometry Analysis
Mass spectrometry analysis is the core step in protein identification. The mass spectrometer separates and detects peptides based on their mass-to-charge ratio (m/z). The steps include:
1. Ionization
Ionize the purified peptides using electrospray ionization (ESI) or matrix-assisted laser desorption/ionization (MALDI) to produce charged peptides.
2. MS1 Analysis
The ionized peptides first enter the primary mass spectrometer for mass-to-charge ratio measurement, resulting in a primary mass spectrum (MS1 spectrum) that shows the distribution of all peptide masses.
3. Selection and Fragmentation
Select specific peptides from the primary spectrum (usually the strongest signals) for fragmentation. The selected peptides are further fragmented into smaller ion fragments in a collision cell using methods such as collision-induced dissociation (CID) or higher-energy collisional dissociation (HCD).
4. MS2 Analysis
The fragmented ions enter the secondary mass spectrometer for mass-to-charge ratio measurement, resulting in a secondary mass spectrum (MS2 spectrum). The MS2 spectrum contains sequence information of the peptides, which is crucial for peptide identification.
Data Processing
Data processing is the process of converting mass spectrometry data into interpretable protein information. It mainly includes the following steps:
1. Data Conversion and Preprocessing
Convert the raw data generated by the mass spectrometer into formats suitable for data analysis software, and perform preprocessing operations such as noise filtering and baseline correction.
2. Peptide Identification
Use database search algorithms (such as SEQUEST, Mascot) to match MS2 spectra with known protein databases, identifying peptide sequences. This process relies on the completeness of the database and the accuracy of the search algorithms.
3. Protein Identification
Assemble the identified peptides into complete protein sequences using bioinformatics tools. Common methods include protein assembly software (such as ProteinPilot, MaxQuant).
4. Data Validation and Quantification
Perform statistical validation of the identification results to ensure their reliability. Additionally, use labeling techniques (such as SILAC, iTRAQ) or label-free quantification methods to quantify proteins.
The workflow of protein identification by tandem mass spectrometry includes three main steps: sample preparation, mass spectrometry analysis, and data processing. Each step requires strict control to ensure high-quality protein identification results. With technological advancements, methods and tools for protein mass spectrometry identification continue to improve, providing strong support for life sciences research.