    How Can the Concentration of a Sample Be Determined Using Gas Chromatography Based on Peak Area and Retention Time?

      Gas chromatography (GC) is a widely employed analytical technique that is used to determine the concentration of a sample based on the retention times and peak areas of its components. The procedure is as follows:

       

      Establishing a Standard Curve

      First, chromatograms for standard solutions of varying concentrations are acquired, with the retention times (tR) and peak areas of each component recorded. A standard curve is then plotted by correlating the concentration of each standard solution with its corresponding peak area. Typically, a linear relationship exists between peak area and concentration. By plotting peak area on the y-axis and concentration on the x-axis, a linear equation, such as y = kx + b, is obtained, where y represents peak area, x represents concentration, and k and b are the slope and intercept of the line, respectively.

       

      Analyzing the Sample

      The sample to be analyzed is then subjected to gas chromatography. Its chromatogram is recorded, and the retention times and peak areas of the sample’s components are noted.

       

      Identifying Components

      The components in the sample can be identified by comparing their retention times with those of the standard solutions on the standard curve.

       

      Calculating Concentration

      The concentration of each component in the sample can be determined by using its peak area and the standard curve equation. This is achieved by substituting the sample’s peak area into the standard curve equation and solving for concentration (x). The formula is as follows:

       

      x = (y - b) / k

       

      Where x is the concentration of the component in the sample, y is its peak area, and k and b are the slope and intercept of the standard curve equation, respectively.

       

      By following these steps, the concentration of each component in the sample can be determined. It is important to note that factors such as instrument calibration and operator technique may influence the results, so it is crucial to ensure accurate operation and instrument precision.

       

      MtoZ Biolabs, an integrated chromatography and mass spectrometry (MS) services provider.

      Related Services

      Quantitative Proteomics Analysis Service

    Submit Inquiry
    Name *
    Email Address *
    Phone Number
    Inquiry Project
    Project Description *

     

    How to order?


    /assets/images/icon/icon-message.png

    Submit Inquiry

    /assets/images/icon/icon-return.png